Sample Requirements


 

  • Must contain viability marker (103Rh, 194Pt, cisPt, or 198Pt).
  • Must be FIXED with minimum 1.6% PFA.
  • Metal Free BSA (Sigma)
  • Must contain Ir191/193 for cell identification.
  • Must be brought in 5mL polypropylene tubes
  • ​MAXIMUM of 1 million pelleted cells per tube
    • Individual or barcoded samples that exceed 1 million cells should be
      split evenly across multiple tubes (for example, a sample with 3 million cells
      should be split into 3 polypropylene tubes with 1 million cells in each tube).
  • Cell Counts (including Cell Viability) obtained prior to final pelleting, must be provided to  staff
  • Always perform 2 final washes using CAS Plus.
  • Samples are added to the Autosampler carousel in Pellet Form ONLY
    • DO NOT ADD BEADS (Autosampler adds EQ6 beads and resuspends the sample in CAS Plus prior to acquisition on-board the instrument.)


 

For novel projects please contact us prior to preparing your samples